Product Details:
|
| Nominal Range: | 0-2000ppm | Maximum Overload: | 5000ppm |
|---|---|---|---|
| Expected Operating Life: | Two Years In Air | Output Signal: | 0.10 ± 0.02μA/ppm |
| Highlight: | 5SF Electrochemical Gas Sensor,Electrochemical Gas Sensor 2000ppm,Sulfur Dioxide Gas Detector |
||
Product Description:
5SF AD526-W00 Sulfur Dioxide Electrochemical Gas Sensor 2000ppm 5 Series
Features:
CiTiceLs may exhibit a response to certain gases in a sample other than the target gas. The table below shows the
typical response of 5SF sensors to a number of common cross-interfering gases. The figures are expressed as a
percentage of the primary sensitivity (i.e. sulphur dioxide = 100%).
Gas Response Gas Response
Carbon monoxide: <3 Hydrogen: <3
Hydrogen sulphide: ≈200 Hydrogen chloride: ≈15
Nitric oxide: 0 Ethylene: <50
Nitrogen dioxide: ≈-125 ** For details of other possible cross-interfering gases contact City Technology
Specifications:
| Temperature Range | -20°C to +50°C |
| Pressure Range | Atmospheric ± 10% |
| Pressure Coefficient | 0.004 % signal/mBar |
| T90 Response Time | <30 seconds |
| Relative Humidity Range | 15 to 90% non-condensing |
| Typical Baseline Range | 0 ± 2ppm equivalent |
![]()
Contact Person: Xu