Product Details:
|
Typical Applications: | Critical Care Anesthesia | Operating Temperature Range: | -20°C To +50°C |
---|---|---|---|
Operating Pressure Range: | 0.5 Bar To 2.0 Bar | Operating Humidity Range: | 0% To 99% RH Non-condensing |
Long-Term Ouput Drift In 100% O2: | < 5% Signal Loss/year | Recommended Storage Temperature: | -10°C To +40°C |
Highlight: | MOX3 Oxygen Gas Sensor,Oxygen Gas Sensor AA829-M10,MOX-3 Oxygen Sensor |
Product Description:
MOX-3 MOX3 AA829-M10 Oxygen Gas sensor Oxygen Battery For Medical Devices
Features:
The purpose of this document is to present the performance specification of the MOX3 oxygen gas sensor.
This document should be used in conjunction with the Operating Principles (OP04) and the Product Safety Datasheet (PSDS 4).
The data provided in this document are valid at 20°C, 50% RH and 1013 mBar for 3 months from the date of sensor manufacture.
Output signal can drift below the lower limit over time. For guidance on the safe use of the sensor, please refer to the Operating Principles (OP04).
CiTiceLs are designed for operation in a wide range of environments and harsh conditions. However, it is important that exposure to high concentrations of solvent vapours is avoided, both during storage, fitting into instruments and operation.
When using sensors with printed circuit boards (PCBs), degreasing agents should be used before the sensor is fitted. Do not glue directly on or near the CiTiceL as the solvent may cause crazing of the plastic.
Intended Use
These sensors are designed to be used to monitor the partial pressure of oxygen in anaesthesia (not including xenon), critical care, neonatal incubators, and general oxygen monitors.
Specifications:
Operating Principle | Partial pressure electrochemical |
Output | 9 mV to 13 mV in 210 mBar O2 |
Measurement Range | 0 mBar to 1500 mBar O2 |
Response Time (T90)* | <15 s (air to 100% O2) |
Baseline Offset* | <200 μV |
Linearity | Linear 0% to 100% O2 |