|
Product Details:
|
Nominal Range: | 0-2000ppm | Maximum Overload: | 4000ppm |
---|---|---|---|
Replaceable Filter Unit: | To Remove Acid Gases | Auxiliary Electrode: | To Compensate For H2 Cross-interference |
Expected Operating Life: | Three Years In Air |
Product Description:
A5F 0-2000PPM Carbon Monoxide CO Sensor
Features:
CiTiceLs may exhibit a response to certain gases in a sample other than the target gas. The table below shows the
typical response of A5F sensors to a number of common cross-interfering gases. The figures are expressed as a
percentage of the primary sensitivity (i.e. carbon monoxide = 100%).
Gas Response Gas Response
Hydrogen sulphide: 0 Hydrogen: <1 (see note)
Sulphur dioxide: 0 Hydrogen chloride: 0
Nitric oxide: 0 Ethylene: ≈35
Nitrogen dioxide: 0 ** For details of other possible cross-interfering gases contact City Technology
Specifications:
Output Signal | 0.075± 0.025 μA/ppm |
Resolution | 1ppm |
Temperature Range | -20°C to +50°C |
Pressure Range | Atmospheric ± 10% |
Pressure Coefficient | 0.02% signal/mBar |
T90 Response Time | <40 seconds |
Contact Person: Xu
Tel: 86+13352990255