Product Details:
|
| Nominal Range: | 0 To 30% O2 | Resolution: | 0.1% O2 |
|---|---|---|---|
| Output Signal: | -85 20 μA In Air | Long Term Output Drift: | < 2% % Signal/ 3 Months |
| Zero Signal: | -10 ~ 5 μA (in N2) | Response Time (T90): | ≤ 15 Seconds |
| Maximum Overload: | 90% O2 | Part Number: | CLE-0231-S00 |
| Highlight: | 30% O2 Range Gas Sensor,O2 Detection Gas Sensor,CLE-0231-S00 Gas Sensor |
||
MINI O2 CLE-0231-S00 Gas Oxygen Sensor For O2 Detection Precision 0 To 30% O2 Range
| Part Number | CLE-0231-S00 |
| Nominal Range | 0 to 30% O2 |
| Maximum Overload | 90% O2 |
| Output Signal | -85 ± 20 μA in air |
| Zero Signal | -10 ~ 5 μA (in N2) |
| Resolution | 0.1% O2 |
| Response Time (T90) | ≤ 15 seconds |
| Long Term Output Drift | < 2% % signal/ 3 months |
| Temperature Range | -20℃ to 50℃ |
| Operating Humidity | 5 to 90%RH non-condensing |
| Pressure Range | 80 to 120 kPa |
| Bias Potential | -600 mV |
| Storage Life | 6 months in sealed container |
| Storage Temperature | 0℃ to 20℃ |
| Expected Operating Life | 2 years in air |
| Warranty | 12 months from date of despatch |
![]()
Contact Person: Xu
Tel: 86+13352990255