Product Details:
|
Nominal Range: | 0-20ppm | Maximum Overload: | 300ppm |
---|---|---|---|
Expected Operating Life: | Two Years In Air | Output Signal: | 1.40 ± 0.30 UA/ppm |
Temperature Range: | -20°C To +50C | Pressure Range: | Pressure Range |
Response Time: | <40 Seconds | Relative Humidity Range: | 15 To 90% Non-condensing |
Highlight: | Nitrogen Dioxide Electrochemical Gas Sensor,3NDH Electrochemical Gas Sensor,0-20 ppm Nitrogen Dioxide Sensor |
3NDH Nitrogen dioxide electrochemical gas sensor 0-20 ppm NO2
The output of a CiTiceL can vary with
temperature. Thegraph hereshowsthe variation
in output with temperature for 3NDH CiTiceLs
based on a sample of about 16 sensors. The
results are shown in the graph as a mean for the
batch, and expressed as a percentage of the
signal at 20°C.
From a statistical viewpoint, for a sample of
this size, the range in values observed for all
sensors of this type will fall within a range three
times the standard deviation above or below the
mean. Assumingtherefore this sample is typical
then the temperature behaviour of all 3NDH
CiTiceLs will fall in the band +3SD to -3SD
Bias Voltage | Not required |
Repeatability | 2% of signal |
Weight | 22g |
Storage Life | Six months in CTL container |
Recommended Storage Temperature | 0-20℃ |
Contact Person: Miss. Xu