Product Details:
|
Operating Principle: | 4-electrode Electrochemical | Measurement Range: | 0 Ppm To 300 Ppm NO |
---|---|---|---|
Maximum Overload: | 1000 Ppm NO | Sensitivity*: | 0.50 MA/ppm ±0.08 MA/ppm |
Filter: | To Remove Effect Of SO2 | ||
Highlight: | Low Range Nitric Oxide Sensor,5NFL Nitric Oxide Sensor |
5NFL Nitric Oxide Sensor Product Code: AF508-W0P Low Range NO CiTicel
Document Purpose
The purpose of this document is to present the performance specification of the 5NFL nitric oxide gas sensor.
This document should be used in conjunction with Operating Principles (OP20) and the Product Safety Datasheet (PSDS 18).
The data provided in this document are valid at 20°C, 50% RH and 1013 mBar for 3 months from the date of sensor manufacture.
Output signal can drift below the lower limit over time. For guidance on the safe use of the sensor, please refer to the Operating Principles (OP20).
Operating Temperature Range (continuous) | -20°C to 40°C [-4°F to 104°F] |
Recommended Storage Temperature | 0°C to 20°C |
Operating Pressure Range | Atmospheric ±10% |
Absolute Pressure Coefficient | <0.01% signal/mbar |
Differential Pressure Coefficient | <0.03% signal/mbar |
Operating Humidity Range | 15%rH to 90%rH non-condensing |
Expected Operating Life | Three years in air |
Long-Term Output Drift | <2% signal loss/month |
Contact Person: Miss. Xu