Product Details:
|
Sensitivity: | -200 To -650 | Response Time: | < 80 |
---|---|---|---|
Zero Current: | < 70 | Noise*: | 15 |
Range: | 20 | ||
Highlight: | High Humidity Nitrogen Dioxide Sensor |
5NO2-B43F Nitrogen Dioxide Sensor 4-Electrode (High Humidity)
LIFETIME:
Zero drift ppb equivalent change/year in lab air 0 to 20
Sensitivity drift % change/year in lab air, monthly test -20 to -40
Operating life months until 50% original signal (24 month warranted) > 24
ENVIRONMENTAL:
Sensitivity @ -20°C (% output @ -20°C/output @ 20°C) @ 2ppm NO2 60 to 80
Sensitivity @ 40°C (% output @ 40°C/output @ 20°C) @ 2ppm NO2 95 to 115
Zero @ -20°C nA 0 to 25
Zero @ 40°C nA -10 to 50
Temperature range | -30 to 40 °C |
Pressure range | 80 to 120kPA |
Humidity range | 30 to 95 % rh continuous |
Storage period | 6 months @ 3 to 20°C (stored in sealed pot) |
Load resistor | 33 to 100 (ISB circuit is recommended) |
Contact Person: Miss. Xu
Tel: 86+13352990255