Maximum Overload: | 1000ppm | Expected Operating Life: | Two Years In Air |
---|---|---|---|
Output Signal: | 0.37 ± 0.07 μA/ppm | Resolution: | 0.25ppm |
Temperature Range: | -40°C To +50°C | Pressure Range: | Atmospheric ± 10% |
Highlight: | 7H Electrochemistry Gas Analyzers,Hydrogen Sulphide Gas Analyzers,200ppm Hydrogen Sulphide Gas Analyzers |
7H Electrochemistry Hydrogen Sulphide For Gas Analyzers 0-200ppm
Cross-sensitivity Data:
CiTiceLs may exhibit a response to certain gases in a sample other than the target gas. 7H CiTiceLs have been tested
with a number of commonly cross-interfering gases and the results are given below. The table shows the typical
response to be expected from a sensor when exposed to a given test gas concentration (relevant to safety, e.g. TLV
levels).
IMPORTANT NOTE:
Connection should be made
via PCB sockets only. Soldering to the pins will
render your warranty void.
Relative Humidity Range | 15 to 90% non-condensing |
Typical Baseline Range (pure air) | -0.6 to +1.9ppm equivalent |
Maximum Zero Shift (+20°C to +40°C) | 2ppm equivalent |
Long Term Output Drift | <2% signal loss/month |
Recommended Load Resistor | 10Ω |
Bias Voltage | Not required (See Application Note #7) |
Repeatability | 1% of signal |
Weight | 12g |