Product Details:
|
Measurement Range: | 0.5 Ppm CO To 1000 Ppm CO (EN 45544 Applications) | Maximum Overload: | 2000 Ppm CO |
---|---|---|---|
Onboard Filter: | To Remove Acid Gases (see Note On Page 3) | Sensitivity*: | 50 NA/ppm ±10 NA/ppm |
T90 Response Time*: | Typically < 20 Seconds | Baseline Offset* (in Clean Air): | < ±3 Ppm CO Equivalent |
Highlight: | Gas Analyzers Gas Sensor |
AB010-R01A-CIT Electrochemical Carbon Monoxide Analogue Gas Sensor For Gas Analyzers
Document Purpose:
The purpose of this document is to present the performance specification of the 1series 1CO carbon monoxide gas sensor.
This document should be used in conjunction with the 1CO Characterisation Note, the Operating Principles (OP08), and the Product Safety Datasheet (PSDS 12.1).
For guidance on sensor performance outside of these limits, please refer to the 1CO Characterisation Note.
Output signal can drift below the lower limit over time. For guidance on the safe use of the sensor, please refer to the Operating Principles (OP08).
Filter Information
Activated carbon cloth filter with high surface area:
• Removes acid gases such as SO2, NO2, and H2S
• 25,000 ppm hours H2S filter capacity
• Protects from exposure to alcohol such as methanol, ethanol, and IPA (<1000 ppm hours)
Poisoning
Gas sensors are designed for operation in a wide range of environments and harsh conditions. However, it is important that exposure to high concentrations of solvent vapours is avoided, both during storage, fitting into instruments, and operation.
When using sensors with printed circuit boards (PCBs), degreasing agents should be used before the sensor is fitted.
Do not glue directly on or near the sensor as the solvent may cause crazing of the plastic.
Recommended Load Resistor | 5 Ω to 10 Ω |
Weight | < 5 g |
Outer Plastic Body Material | Modified PPO |
O-ring Material | FKM75 ±5 shore A |
Contact Material | Gold plated |
Operating Temperature Range | -40°C to +60°C |
Operating Pressure Range | 600 mbar to 1200 mbar |
Long Term Output Drift* | < 5% signal loss per annum |