Product Details:
|
| Sensitivity: | 160 NA/ppm ± 40 NA/ppm | Zero Current: | ± 20 NA |
|---|---|---|---|
| Maximum Overload: | 50 Ppm | Resolution (16Bit ADC): | 0.01 Ppm |
| Response Time: | T50 < 10s, T90 < 30s | Repeatability: | 1% |
| Highlight: | Industrial Safety Gas Sensor,High Selectivity Gas Sensor,Solid Polymer Electrochemical Gas Sensor |
||
ES4-H2S-10 ES4-H2S-50 Solid Polymer Electrochemical Gas Sensing Technology Hydrogen Sulfide Gas Sensor Detects With High Selectivity A Wide Variety Of Gases For Industrial Safety
Features:
Detects with high selectivity a wide variety of gases
Long lifetime > 3 years
No-poisoning
Typical warm-up time in seconds
Fast response time
nA power consumption
Linear output
No zero line drift
Better signal to noise ratio
Wide temperature range of -40°C to +55°C
Excellent sensitivity at low temperatures
No leakage
Small size
RoHS compliant
| Operating Temperature Range | -40℃ to +55℃ |
| Operating Humidity Range | 15-95 %RH. Non-condensing |
| Operating Pressure Range | 800 to 1200 hPa |
| Storage Temperature | 0 to 20℃(Optimum temp. 4 to 6℃ ) |
| Operating Principle | Amperometric, 3-electrode |
![]()
Contact Person: Miss. Xu