Operating Principle: | Partial Pressure Electrochemical | Output: | 9 MV To 13 MV In 210 MBar O2 |
---|---|---|---|
Measurement Range: | 0 MBar To 1500 MBar O2 | Response Time (T90): | <15 S (air To 100% O2) |
Baseline Offset: | <200 μV | Linearity: | Linear 0% To 100% O2 |
Highlight: | Widely Used Oxygen Gas Sensor,MOX4 Oxygen Gas Sensor |
MOX4 Oxygen Sensor
Part Number: AA829-M20
Document Purpose:
The purpose of this document is to present the performance specification of the MOX4 oxygen gas sensor.
This document should be used in conjunction with the Operating Principles (OP04) and the Product Safety Datasheet (PSDS 4).
The data provided in this document are valid at 20°C, 50% RH and 1013 mBar for 3 months from the date of sensor manufacture.
Output signal can drift below the lower limit over time. For guidance on the safe use of the sensor, please refer to the Operating Principles (OP04).
If the Sensor is Dropped:
If a sensor is dropped, then it should be placed in quarantine for 24 hours and a follow-up check made by a 2-point calibration.
Mechanical Installation:
When installing the sensor, it must only be screwed in hand-tight and a gas tight seal ensured. Spanners and similar mechanical aids may not be used, as excessive force may damage the sensor thread.
RFI/EMI Susceptibility:
MediceLs contain metal and may be susceptible to RFI or EMI. They are not suitable for use in MRI environments. For further information, please contact City Technology.
Housing Material | White ABS |
Weight | 42 g (nominal) |
Orientation | Any |
Typical Applications | Critical care anesthesia |