Measuring Range: | 0 - 100 Ppm | Sensitivity: | 275 - 500 NA/ppm |
---|---|---|---|
Response Time (t90): | < 60 S | Zero Current: | - 100 - + 100 NA |
Resolution: | 5 Ppb | Maximum Overload: | 200 Ppm |
Highlight: | SO2-B4 sulfur dioxide gas sensor,electrochemical gas sensor for air quality,flue gas analysis SO2 sensor |
SO2-B4 Electrochemical Sulfur Dioxide Gas Sensor For Air Quality Monitoring Flue Gas Analysis
Technical Features:
Application Scenarios:
It is suitable for fixed - installation sensing heads, portable safety instruments, urban/indoor air monitoring and stack gas analyzers, etc., and can provide reliable data support for air quality monitoring and industrial emission detection.
Measuring Range | 100 ppm |
Sensitivity | 275 - 475 nA/ppm (at 2 ppm SO2) |
Response Time (t90) | < 30 s (from zero to 2 ppm SO2) |
Zero Current in Zero Air at 20℃ | - |
Noise (±2 standard deviations, ppb equivalent) | 5 |
Resolution | < 1 ppm |
Linearity | 0 - 2 ppb error at 100 ppm SO2, linear at 0 and 10 ppm SO2 |
Over - gas Limit | 200 ppm |
Zero Drift (ppb equivalent change/year in lab air) | < ±20 |
Sensitivity Drift (% change/year in lab air, monthly test) | < ±15 |
Operating Life | > 36 months (until 50% of the original signal) |
Sensitivity at - 20℃ | 70 - 82% (output at - 20℃ / output at 20℃ @ 2 ppm SO2) |
Sensitivity at 50℃ | 95 - 110% (output at 50℃ / output at 20℃ @ 2 ppm SO2) |
Zero at - 20℃ (nA change from 20℃) | 0 - 10 |
Zero at 50℃ (nA change from 20℃) | 10 - 30 |
Filter Capacity | 450 ppm·hrs |
H2S Sensitivity (% measured gas @ 5 ppm H2S) | < 2 |
NO2 Sensitivity (% measured gas @ 5 ppm NO2) | < - 160 |
Cl2 Sensitivity (% measured gas @ 5 ppm Cl2) | < - 40 |
NO Sensitivity (% measured gas @ 5 ppm NO) | < - 2 |
CO Sensitivity (% measured gas @ 5 ppm CO) | < 2 |
H2 Sensitivity (% measured gas @ 100 ppm H2) | < 0.5 |
C2H4 Sensitivity (% measured gas @ 100 ppm C2H4) | < 1 |
NH3 Sensitivity (% measured gas @ 20 ppm NH3) | < 0.1 |
CO2 Sensitivity (% measured gas @ 5% CO2) | < 0.1 |
Temperature Range | - 30 to 50℃ |
Pressure Range | 80 - 120 kPa |
Humidity Range | 15 - 90% RH (continuous) |
Storage Period | 6 months (@3 - 20℃, stored in sealed pot) |
Load Resistor | 33 - 100 Ω |
Weight | < 13 g |