Product Details:
|
| Oxygen Concentration Measurement Range: | 0.1%–100% Vol O2 | Oxygen Partial Pressure Range: | 2–3000 Mbar |
|---|---|---|---|
| Accuracy After Calibration: | ±0.5% O2 | Resolution: | 0.01% O2 |
| Highlight: | oxygen sensor probe rapid response,analog output oxygen sensor,gas sensor with warranty |
||
O2S-FR-T2-18BM-C Oxygen Sensor Probe Rapid Response Type Analog output
Operating Principles and Features (Distinguishing it from the 4OX-V electrochemical oxygen sensor)
1. Uses a non-consumable zirconia element with an exceptionally long service life (5–10 years); unlike electrochemical sensors, it does not fail due to electrolyte dry-out.
2. Requires no reference gas; a single-point calibration in fresh air (20.7% O₂) is sufficient.
3. Immune to interference from temperature, humidity, or minor fluctuations in atmospheric pressure; offers outstanding stability in harsh flue gas environments.
4. Relies on a built-in high-frequency oxygen pump to cyclically draw gas into and out of the chamber, calculating oxygen concentration based on pump cycles; delivers excellent linearity.
Specification:
| Operating voltage | 4.35 VDC ±0.1 V; current: 1.85 A |
| Standby/thermal maintenance | 2 VDC, 0.85 A (intermittent power-saving mode) |
| Pump body impedance at 700°C | <6 kΩ |
| Gas flow velocity tolerance | 0–10 m/s |
![]()