Product Details:
|
| Minimum Detection Level: | 100 Ppb | Operating Range: | 0-4000PPM |
|---|---|---|---|
| Response Time (T90): | 3.5 Seconds | Warm Up Time: | 5 Seconds |
| Full Stabilisation Time: | 5 Minutes | Offset Voltage: | 40-75 MV |
| Highlight: | PID sensor for petrochemical processing,photoionisation detector for industrial hygiene,PID gas sensor for chemical processing |
||
PIDX-A-04K Photoionisation Detector (PID) Sensor for Petrochemical & Chemical Processing Industrial Hygiene
Dual-range performance 0-4,000 ppm operating range with a linear range of 0-400 ppm, combining wide coverage with reliable low-concentration measurement.
High sensitivity & low detection limit Typical sensitivity of ~1.8 mV/ppm at 100 ppm and a minimum detection level of 100 ppb.
Intrinsically safe (IS) design Certified Ex ia IIC Ga, requiring no additional circuits or external fuses for safe operation in hazardous areas.
Enhanced repeatability & stability The PIDX platform delivers improved sensor-to-sensor repeatability and a stable baseline over a wide temperature and humidity range, reducing calibration overhead and improving long-term trend accuracy.
Low power consumption Approximately 92 mW, helping extend battery life in portable and battery-powered instruments.
User-replaceable lamp & grid The 10.6 eV lamp and electrode grid are field-replaceable, with a typical lamp life of 10,000 hours.
Fast response T90 response time of 3.5 seconds, with a 5-second warm-up time.
Rugged construction Weighs <12 g, with an on-board filter to remove liquids and particulates.
The PIDX-A-04K is engineered for VOC safety monitoring in the following sectors:
Oil & gas Fugitive emission detection, process area monitoring, and leak screening in upstream, midstream, and downstream operations.
Petrochemical & chemical processing Solvent handling, reactor area monitoring, and confined space entry.
Industrial hygiene Worker exposure assessment in painting, coating, printing, and other VOC-intensive processes.
Marine & container coating Stable VOC detection under the complex conditions of shipbuilding and container painting operations.
Spray booths & open workshops Adaptable to both enclosed spray rooms and open production areas for continuous VOC surveillance.
Fence-line & ambient monitoring Medium-to-high concentration VOC leak points and fugitive emissions at industrial perimeters.
| Category | Parameter | Specification |
|---|---|---|
| Performance | Target Gas | VOCs with ionisation potential < 10.6 eV |
| Minimum Detection Level | 100 ppb | |
| Linear Range | 0–400 ppm | |
| Operating Range | 0–4,000 ppm | |
| Typical Sensitivity | 1.8 mV/ppm ± 10% (tested at 100 ppm) | |
| Response Time (T90) | 3.5 seconds | |
| Warm-Up Time | 5 seconds | |
| Full Stabilisation Time | 5 minutes | |
| Offset Voltage | 40–75 mV | |
| Electrical | Power Consumption | 92 mW |
| Supply Voltage | 3.2–5.5 VDC | |
| Output Signal | 0.040–2.85 V | |
| Environmental | Operating Temperature | −20 °C to +60 °C (intrinsically safe); −40 °C to +65 °C (operating) |
| Relative Humidity | 0–95% RH, non-condensing | |
| Humidity Sensitivity | Near zero (0–95% RH) | |
| Mechanical & Life | Weight | < 12 g |
| Expected Sensor Life | 5 years (excluding replaceable lamp and grid) | |
| Lamp Life | 10,000 hours (typical, user-replaceable) | |
| On-Board Filter | Removes liquids and particulates | |
| Position Sensitivity | None | |
| Certifications | Intrinsic Safety (IS) | Ex ia IIC Ga; ATEX, UKEX, IECEx certified; no additional circuits or external fuses required |
| Warranty | Electronics & Housing | 24 months |
| Lamp | 12 months |
![]()
Contact Person: Miss. Xu
Tel: 86+13352990255